https://cdn.ampproject.org/v0/amp-auto-ads-0.1.js https://pagead2.googlesyndication.com/pagead/js/adsbygoogle.js?client=ca-pub-8347311151393716 Dr. Yousef for Lab Information

Sunday, November 2, 2025

CERULOPLASMIN Test. lab information about test and diagnosis of diseases

______________________________________________________________________________

Ceruloplasmin Test. Lab information about test and diagnosis of diseases. 


Ceruloplasmin Test – Manual of Medical Laboratory Techniques

Purpose:
To measure the amount of ceruloplasmin in human serum. Low levels are reported in Wilson’s disease and nephrotic syndrome, and it is valuable in differentiating chronic liver diseases from Wilson’s disease.

Principle:
Ceruloplasmin, a ferro-oxidase enzyme, catalyzes the oxidation of certain polyamines. Its activity on p-phenylenediamine is measured to determine the amount of ceruloplasmin in serum.

Performance Specifications:

  • Linearity: up to 60 mg/dL
  • Measurement range: 16–60 mg/dL
  • Sensitivity: 16 mg/dL

Primary Sample:

  • Use serum as the specimen.
  • Collect 4 mL of venous blood in a plain red vacutainer.
  • Allow to clot for 30 minutes, then centrifuge at 2500 rpm for 10 minutes.
  • Avoid hemolyzed or turbid samples.
  • Analyze within 3 hours of collection, or store serum at –20°C for up to 7 days.

Reagents and Consumables:

  • p-Phenylenediamine hydrochloride: 5 g in 1 L distilled water (purified and crystallized).
  • Acetic acid (1 M): 60 mL glacial acetic acid diluted to 1 L.
  • Sodium acetate (1 M): 136 g per liter of water.
  • Acetate buffer (400 mM, pH 5.5): Prepared by mixing acetic acid and sodium acetate.
  • Sodium azide or sodium fluoride: Preservatives (5 g/L or 20 g/L).

Instrument:
Spectrophotometer (read at 530 nm).

Procedure:






Calculation:


Reference Range:

  • Normal: 20 – 40 mg/dL

Critical/Alert Values:

  • Not applicable.

Potential Sources of Error:

  • Hemolyzed serum may give falsely elevated values.
  • Samples not analyzed promptly must be stored at –20°C to maintain accuracy.

References:
Ramakrishnan, Manual of Medical Laboratory Techniques, pp. 25–27.
King, J. Practical Clinical Enzymology, 1965.


Produce by : Dr. Yousef AL-Adbai. 

ALBUMIN Test. lab information about test and diagnosis of diseases.

_______________________________________

ALBUMIN Test. Lab information about test and diagnosis of diseases. 


Albumin Test – Manual of Medical Laboratory Techniques

Purpose:
Quantitative estimation of albumin in human serum by photometric method using bromocresol green (BCG) dye binding. Elevated serum albumin levels are associated with dehydration. Low serum albumin levels indicate potential malnutrition, liver diseases, kidney disorders (especially nephrotic syndrome), and rheumatoid arthritis.

Principle:
Albumin acts as a cation at pH 3.8 and selectively binds to the anionic dye bromocresol green, forming a green-colored complex. The intensity of the color, measured at 630 nm, is directly proportional to the concentration of albumin in the sample.

Performance Specifications:

  • Linearity: up to 6.0 g/dL
  • Measurement range: 0.5–6.0 g/dL
  • Sensitivity: 0.5 g/dL
  • Specificity: certain drugs (e.g., Ampicillin) may interfere with dye-binding properties. Only human albumin standards should be used.

Primary Sample:

  • Use serum (fasting sample recommended).
  • Avoid venostasis during collection to prevent hemoconcentration.
  • Collect 4 mL venous blood in a plain red vacutainer; allow to clot for 30 minutes.
  • Centrifuge at 2500 rpm for 10 minutes.
  • Avoid hemolyzed or turbid samples.
  • Analyze within 3 hours or store serum at 2–8 °C (up to 30 days).

Reagents and Equipment:

  • Reagent: Bromocresol Green (BCG) buffer, pH 3.68
  • Standard: Bovine albumin fraction V (5 g/dL)
  • Instrument: Semi-auto analyzer (e.g., RA 50)
  • Wavelength: 628 nm
  • Temperature: 37 °C
  • Incubation time: 10 minutes

Procedure:

  1. Run reagent blank with distilled water.
  2. Process standard, then patient samples.
  3. Measure absorbance at 628 nm.

Calculation:


Reference Range:

  • Adults: 3.5 – 5.0 g/dL

Critical Values:

  • Below 2.0 g/dL indicates severe hypoalbuminemia.

Interpretation:

  • Low albumin: liver disease, nephrotic syndrome, malnutrition, chronic illness, or pregnancy.
  • High albumin: dehydration.

Potential Sources of Error:

  • Hemolyzed serum may give false high values.
  • Non-human albumin standards or contaminated reagents may cause inaccurate results.

________________________________________

References:
Ramakrishnan, Manual of Medical Laboratory Techniques, pp. 23–25.

Produce by :Dr. Yousef AL-Adbai

______________THANK_YOU_____________

Saturday, November 1, 2025

TOTAL Protein. Information about test and diagnosis of diseases

                     TOTAL PROTEINS.
Information about test and diagnosis of diseases 

1. Purpose:
Estimation of total protein in serum/body fluids by Biuret method. Low protein levels are observed in malnutrition, acute or chronic liver diseases, nephrotic syndrome, water intoxication, salt retention syndromes, and massive intravenous infusions. Elevated protein levels are observed in dehydration due to vomiting, diarrhea, Addison’s disease and diabetic ketoacidosis. High protein levels of over 2 g/dL in body fluids are suggestive of inflammation or malignancy and are called exudates. 
_______________________________________
2. Principle: 
Peptide bonds of proteins in serum react with cupric ions in alkaline solutions to form a blue colored complex, the absorbance of which is measured at 578 nm. The intensity of the blue color is proportional to the amount of protein present. The reaction sequence employed in the assay of total proteins is as follows:
_______________________________________
3. Performance specifications: 
3.1. Linearity: Up to 12 g/dL 
3.2. Measurement range: This method has a measurement range of 5.3–8.4 g/dL of total protein in serum and body fluids.
3.3. Sensitivity: The minimum detection limit by the kit is 5.3 g/dL. 
_______________________________________
4. Primary sample:
4.1. Use serum/body fluids (Pleural, Pericardial, Ascitic fluid) as specimen for the test. 
4.2. Collect blood sample in a red color vacutainer tube, separate serum within 30 minutes of collection. 
4.3. Process the sample on the same day within 1 hour of collection. If analysis is done on the next day, separate the serum and store it at 2–8°C for up to 30 days.
_______________________________________
5. Type of container and additive:
Use plain vacutainer tubes for collecting samples. No additive/Preservative is needed to be added.
_______________________________________
6. Reagents/Consumables: 
6.1. Biuret reagent. 
6.2. Total protein standard.
7. Instrument: Semi-autoanalyzer
_______________________________________
8. Procedure:
8.1 Switch on the machine and press “FLUSH “button by keeping the tubing in a container with 2% detergent for 2 minutes followed by distilled water for 2 minutes. 
8.2. Press “PROC”. Different test procedures will be displayed. 
8.3. Select the test to be processed by entering its number and then press “ENTER” key. 
8.4. Now the assay parameters of the specific test procedure will be displayed. Note down the volume of the reagent and the sample to be used. 
8.5. Feed the reagent blank with each batch of patient samples and ensure the absorbance of the blank is less than 0.150, if the absorbance of the ‘blank is more than 0.150’ discard the reagent at 546 nm. 
8.6. Then feed the test samples and record the values. 
8.7. Check whether the sample is hemolyzed or icteric before processing. If the sample is lysed, collect another sample and proceed. If it is icteric or lipemic, dilute the sample 1 in 10 with distilled water and proceed. Multiply the result displayed by dilution factor 10.
_______________________________________
Assay type: End point,
Wavelength: 546 nm.
Sample volume: 10 μL, 
Reagent volume: 1000 μL
Incubation time: 20 min at RT, Temperature: 37oC
_______________Dr_Yousef______________
9. Interferences:
9.1. Feed the reagent blank with each batch of patient samples and ensure the absorbance of the blank is less than 0.150, If the absorbance of the ‘blank is more than 0.150’ discard the reagent at 546 nm. 
9.2. Keep the reconstituted reagent at 2–8ºC. Discard the same if it develops precipitate. 
9.3. Highly hemolytic or icteric samples, prepare sample blank by adding 1 mL of 0–9% saline to 10 microliter samples. The value of the blank is subtracted from the corresponding sample value.
_______________________________________
10. Calculating results:
11. Biological reference range:
Adults: 6.6–8.4 g/dl.
12. Critical/Alert values:
Below 5.0 g/dL and above 9.0 g/dl.
13. Laboratory interpretation: 
Increase of proteins in dehydration, multiple myeloma and chronic infections (gammopathy); 
decrease in malnutrition, liver diseases, nephrotic syndrome.
14. Potential sources of variability: 
14.1. The reagent is linear to 12.0 g/dL. Samples with values above 10 g/dL should be diluted 1:1 with 0.9% saline, re-run, and the result multiplied by two (2) 
14.2 The biuret procedure is not sensitive at low ranges (< 1 g/dL). Do not use for urine or spinal fluid.
================================
BIBLIOGRAPHY
1. Henry J, Winkelman JW. Clinical Chemistry Principles and Technique. Harper and Row, 2nd edn 1974. 
2. Stricklad RD, Freeman ML, Gurule FF. Copper Binding by proteins in alkaline solution. Anal Chem 1961;33.

Reference: Manual of Medical Laboratory Techniques

Produce by: Dr. Yousef Al Adbai
Laboratorist. 
______________THANK_YOU______________


triacylglycerols (triglycerides). Information about test and diagnosis of diseases

TRIACYLGLYCEROLS (TRIGLYCERIDES)
Information about test and interpretation about Result 

1. Purpose: 
Quantitative estimation of triacylglycerols in human serum by enzymatic method using Glycerol -3 Phosphate Oxidase (GPO) Measurement of triglycerides in conjunction with other lipid assays is used in screening the lipid status of an individual to detect atherosclerotic risks and in monitoring the response to lipid lowering measures triglyceride determinations when performed are useful in the diagnosis of primary and secondary hyperlipoproteinemia. They are also of interest in following the course of diabetes mellitus, nephrotic syndrome, biliary obstruction and various metabolic abnormalities due to endocrine disturbances.
 2. Principle:
 The procedure involves hydrolysis of triglycerides by lipoprotein lipase. The glycerol concentration is then determined by enzymatic assay coupled with Trinder reaction that terminates in the formation of a quinoneimine dye which is generated from 4-aminoantipyrine and 4-chlorophenol by hydrogen peroxide under the catalytic action of peroxidase. The amount of the dye formed, determined by its absorption at 500 nm, is directly proportional to the concentration of triglycerides in the sample.
3. Performance specifications:
3.1. Linearity: Up to 1000 mg/dL of serum 
3.2. Measurement range: 1–1000 mg/dL of cholesterol in serum
3.3. Sensitivity: The minimum detection limit by this kit is 1 mg/dL.
4. Primary sample:
4.1. Use only fasting serum as specimen 
4.2. Collect blood sample after an overnight fast of 12–14 hours when testing is a part of lipid profile 
4.3. Collect 4 mL of venous blood in a plain vacutainer tube. 
4.4. Allow the tube to stand for 30 minutes and separate the serum by centrifugation at 2500–3000 rpm for 5–10 minutes 
4.5. Do not use lysed serum for testing as it may give very high results 
4.6. Do not use contaminated/turbid samples for testing 
4.7. Process the sample on the same day within 3 hours of collection. 
4.8. If analysis is not done on the same day/within 3 hours of collection, separate the serum and store it at 20–25 °C for up to 2 days or at 4–8 °C for up to 7 days
5. Type of container and additive:
Use plain vacutainer tubes for collecting samples. No additive/preservative is needed to be added.
6. Reagents/Consumables:
Lipoprotein lipase, magnesium acetate, 4 aminoantipyrine, glycerol-3-phosphate oxidase, glycerol kinase, peroxidase, triglyceride standard 200 mg/dL triglycerides as triolein. 
7. Instrument: Semi-autoanalyzer.
8. Procedure:
8.1. Switch on the machine and press “FLUSH “button by keeping the tubing in a container with 2% detergent for 2 minutes followed by distilled water for 2 minutes. 
8.2. Press “PROC”. Different test procedures will be displayed. 
8.3. Select the test to be processed by entering its number and then press “ENTER” key. 
8.4. Now the assay parameters of the specific test procedure will be displayed. Note down the volume of the reagent and the sample to be used. 
8.5. Feed the reagent blank with each batch of patient samples and ensure the absorbance of the blank is less than 0.300 at 520 nm if the absorbance of the ‘blank is more than 0.300, discard the reagent. 
8.6. Then feed the test samples and record the values. 
8.7. Check whether the sample is hemolyzed, icteric before processing. If the sample is lysed, collect another sample and proceed. If it is icteric or highly lipemic, dilute the sample 1 in 10 with distilled water and proceed. Multiply the result displayed by dilution factor 10.
_______________________________________
Assay: End point.    
Reagent volume: 1000 μL
Wavelength: 546 nm
Sample volume: 10 μL
Temperature: 30°C
Zero setting with distilled water
Incubation time: 5 minutes
Conc. of standard: 200 mg/dL
_______________________________________
9. Calculating results:
10. Biological reference range:
Male: 60–165 mg/dL 
Female: 40–140 mg/dL
11. Critical/Alert level values: 400 mg/dL
12. Laboratory interpretation: Triacylglycerolemia is a risk factor for myocardial infarction.
 TGA is phenomenally increased in an eye disease lipemic retinalis.
13. Potential sources of variability: 
13.1. Lysed serum specimens may give falsely elevated values 
13.2. Do not use if the reagent is turbid as it indicates contamination of the reagent and if the absorbance of the blank reagent is more than 0.300.
================================
BIBLIOGRAPHY:
1. Annoni G, Bottasso BM. Ciaci D, Donato MF Tripoli A, Lab JJ. Res Lab Med 1982;9:115. 
2. Buccolo G, David M. Clin. Chem 1973;19:476. 
3. Werner M, Gabrielson DG, Estman G. Clin Chem 1981;21:268.

Reference: Manual of Medical Laboratory Techniques
Produce by: laboratorist
Dr. Yousef Al Adbai 

Tuesday, October 28, 2025

LOW-DENSITY LIPOPROTEIN (LDL) CHOLESTEROL. Information about test and diagnosis of diseases

LDL (Low-Density Lipoprotein) Cholesterol — Calculation, Purpose & Reference Range

LOW-DENSITY LIPOPROTEIN (LDL) CHOLESTEROL

LDL cholesterol (LDL-C) is commonly known as the “bad cholesterol.” It plays a key role in assessing cardiovascular risk and is calculated from other lipid profile components.

1. Purpose

The estimation of LDL cholesterol in human serum is performed by calculation using a standard formula. Measuring serum LDL cholesterol helps in:

  • Screening lipid status to detect atherosclerotic risks
  • Monitoring response to lipid-lowering measures
  • Diagnosis and classification of hyperlipidemias

A strong relationship exists between serum LDL cholesterol and the risk of coronary heart disease.
LDL > 130 mg/dL is considered a risk factor for coronary and cerebrovascular disease and is also useful for lipoprotein phenotyping.

2. Principle

The LDL cholesterol concentration is determined by calculation rather than direct chemical assay. It uses the **Friedewald Formula**, which estimates LDL cholesterol from total cholesterol, triglycerides, and HDL levels.

3. Procedure / Calculation

The LDL cholesterol (mg/dL) is calculated as:

LDL = Total Cholesterol − (Triglycerides / 5 + HDL)

⚠️ Valid only when triglyceride level is below 400 mg/dL.

4. Reference Range

ClassificationLDL Cholesterol (mg/dL)
Desirable< 130
Borderline elevation130 – 159
High / Elevated> 160

Calculation valid up to triglyceride levels of 400 mg/dL.

5. Clinical Interpretation

  • Optimal LDL (< 100 mg/dL): Minimal atherosclerotic risk
  • Borderline (130–159 mg/dL): Moderate risk — lifestyle modification advised
  • High (> 160 mg/dL): Significant cardiovascular risk — medical intervention recommended
Critical value: LDL cholesterol above 190 mg/dL indicates severe hyperlipidemia and a high risk of coronary artery disease.
LDL cholesterol calculation is based on the Friedewald formula. For accurate assessment, use fasting samples and correlate with clinical findings.

HIGH-DENSITY LIPOPROTEIN - (HDL) CHOLESTEROL. Information about test and diagnosis of diseases

HDL (High-Density Lipoprotein) Cholesterol — Test Procedure & Interpretation

HIGH-DENSITY LIPOPROTEIN — (HDL) CHOLESTEROL

Quantitative estimation of HDL cholesterol in human serum by magnesium-phosphotungstate precipitation followed by enzymatic cholesterol assay.

1. Purpose

Quantitative estimation of HDL cholesterol in human serum by precipitation method — precipitation of VLDL and LDL (using magnesium ions and phosphotungstic acid), followed by estimation of HDL cholesterol using the cholesterol esterase–cholesterol oxidase method.

Measurement of serum HDL cholesterol is useful to screen lipid status, detect atherosclerotic risk, monitor response to lipid-lowering measures, and classify hyperlipidemias. An inverse relationship exists between serum HDL cholesterol and coronary heart disease risk. HDL < 30 mg/dL is considered a risk factor for coronary and cerebrovascular disease.

2. Principle

Phosphotungstate / Mg2+ precipitate VLDL, LDL and chylomicrons; HDL remains in the supernatant. After centrifugation the supernatant (HDL fraction) is assayed for cholesterol using the cholesterol esterase–cholesterol oxidase enzymatic method.

Schematic:
Serum/Plasma —(Phosphotungstate / Mg2+)→ HDL (supernatant) + (LDL + VLDL + CM in precipitate)

3. Performance specifications

CharacteristicValue
LinearityUp to 400 mg/dL of serum
Measurement range1 – 400 mg/dL
SensitivityMinimum detection limit: 1 mg/dL
SpecificityCholesterol oxidase oxidizes some cholesterol analogs (e.g., dihydrocholesterol, 7-dehydrocholesterol) — these analogs are not normally present in appreciable amounts in serum.

4. Primary sample

  • Use only fasting serum as specimen.
  • Collect 4 mL venous blood in a heparin vacutainer.
  • Allow to stand 30 minutes; separate serum by centrifugation at 2500–3000 rpm for 5–10 minutes.
  • Avoid icteric/hemolyzed or turbid/contaminated samples — they may give falsely high results.
  • Process sample the same day within 3 hours of collection. If delayed, separate serum and store at 2–8 °C for up to 7 days.

5. Container & additive

Heparin vacutainer. No preservatives required.

6. Reagents / Consumables

  • Precipitating reagent: phosphotungstic acid (2.4 mM) + magnesium chloride (40 mM)
  • Cholesterol reagent components: 4-aminophenazone, phenol
  • Enzymes: cholesterol esterase, cholesterol oxidase, horseradish peroxidase
  • Buffer (pH 6.8), stabilizers, and fillers
  • HDL-cholesterol standard: 50 mg/dL

7. Instrument

Semi-automatic analyzer (follow instrument specific warm-up and maintenance procedures).

8. Procedure

  1. Bring reagents to room temperature.
  2. Add 500 µL serum + 500 µL HDL precipitating reagent. Mix and centrifuge at 4000 rpm for 10 min to obtain a clear supernatant.
  3. Assay the supernatant for HDL cholesterol using the cholesterol reagent (same steps as total cholesterol assay).
  4. Instrument prep: flush tubing with 2% detergent for 2 minutes, then distilled water for 2 minutes.
  5. Select Absorbance mode, choose the test, note reagent and sample volumes.
  6. Run a reagent blank with each batch; blank absorbance must be < 0.100. Discard reagent if blank >= 0.100.
  7. Feed patient samples and record values.
  8. Check samples for hemolysis or icterus. If hemolyzed — recollect. If highly icteric or lipemic, dilute 1:10 with distilled water and multiply result by 10.
Assay details (end point):
Reagent volume: 1000 µL  •  Sample volume: 50 µL  •  Wavelength: 510 nm  •  Temperature: 37 °C
Zero: reagent blank  •  Incubation: 10 minutes  •  Standard concentration: 50 mg/dL

9. Interferences & precautions

  • Avoid blood collection in fed state — use fasting samples.
  • Do not use samples kept above 2–8 °C for ≥1 day.
  • Hemolyzed samples may give falsely elevated values; turbid reagent or high reagent blank indicates contamination — do not use.

10. Calculations

Calculate HDL cholesterol from absorbances using the standard:

Sample concentration = (Sample absorbance / Standard absorbance) × Concentration of standard

11. Biological reference range

Normal HDL cholesterol: 30 – 60 mg/dL

12. Alert / Critical values

HDL cholesterol < 30 mg/dL — considered a risk factor for coronary and cerebral vascular disease.

13. Laboratory interpretation

  • HDL/Total cholesterol ratio < 0.2 indicates increased coronary heart disease risk.
  • Equivalently, a Total cholesterol / HDL cholesterol ratio ≈ 5 is considered unfavorable.

14. Potential sources of variability

  • Hemolysis → falsely elevated values.
  • Turbid reagents or high reagent blank (≥ 0.100) → reagent contamination; do not use.
Bibliography
  1. Castelli WP. Circulation 1977;55:767.
  2. Castelli WP. Metabolic Therapy 1977;6:1.
  3. Gordon T, et al. Am J Med 1977;62:707.
Prepared as a laboratory test protocol for educational and clinical reference. Always correlate results clinically and follow local laboratory regulations and instrument manufacturer instructions.

CHOLESTEROL TEST. Information about test and diagnosis of diseases

 

CHOLESTEROL TEST (CHOD-PAP METHOD)

1. Purpose

Quantitative estimation of total cholesterol in human serum by CHOD-PAP method (enzymatic photometric method). Measurement of serum cholesterol is useful in screening the lipid status of individuals to detect atherosclerotic risks and in monitoring the response to lipid-lowering measures. It is also useful in the diagnosis and classification of hyperlipidemias. Other conditions such as hepatic and thyroid diseases also influence cholesterol levels.

2. Principle

Cholesterol esters are hydrolyzed by cholesterol esterase to produce free cholesterol and fatty acids. Hydrogen peroxide is produced from the oxidation of cholesterol by cholesterol oxidase. In a coupled reaction catalyzed by peroxidase (POD), red quinoneimine dye is formed from 4-aminoantipyrine, phenol and hydrogen peroxide. The absorbance of this dye at 500 nm is proportional to the cholesterol concentration in the sample (Trinder’s reaction).

Cholesterol ester --(Chol. esterase)--> Cholesterol + Fatty acids
Cholesterol --(Chol. oxidase)--> 2 H₂O₂ + Cholesten-4-en-3-one
2 H₂O₂ + 4-Aminoantipyrine + Phenol --(POD)--> Red quinoneimine + H₂O (Red dye)
  

3. Performance Specifications

  • Linearity: Up to 1000 mg/dL of serum
  • Measurement range: 1 – 1000 mg/dL
  • Sensitivity: Minimum detection limit = 1 mg/dL
  • Specificity: Cholesterol oxidase is not totally specific; analogs may also be oxidized but occur rarely in serum.

4. Primary Sample

  1. Use only plasma as specimen.
  2. Collect 4 mL venous blood in a heparin vacutainer tube.
  3. Centrifuge at 2500 rpm for 10 minutes.
  4. Do not use lysed or contaminated samples.
  5. Process within 3 hours of collection; otherwise recollect sample.

5. Container and Additive

Use plain or heparin vacutainer tubes for sample collection.

6. Instrument

Semi-auto analyzer.

7. Reagents / Consumables

  • 4-Aminoantipyrine
  • Phenol
  • Cholesterol esterase
  • Cholesterol oxidase buffer (pH 6.8)
  • Cholesterol standard (200 mg/dL in alcohol)

8. Procedure

  1. Switch on the analyzer and flush tubing with 2% detergent for 2 min, then with distilled water for 2 min.
  2. Select “PROC”, choose test number, and press ENTER.
  3. Check reagent blank (absorbance < 0.300). If higher, discard reagent.
  4. Run test samples and record readings.
  5. If sample is icteric or lipemic, dilute 1:10 with distilled water and multiply result × 10.

Assay Conditions:

  • Type: End-point
  • Wavelength: 510 nm
  • Temperature: 37 °C
  • Incubation time: 5 min
  • Reagent volume: 1000 µL
  • Sample volume: 10 µL
  • Standard concentration: 200 mg/dL

9. Interference

Avoid hemolyzed or icteric samples. Over-time reagent discoloration (light pink) is acceptable, but discard if absorbance > 0.3 OD vs distilled water at 510 nm.

10. Calculation of Results

Sample Concentration (mg/dL) = (Sample Absorbance × Standard Concentration) / Standard Absorbance

11. Biological Reference Range

  • Serum: 135 – 220 mg/dL
  • Desirable: ≤ 200 mg/dL
  • Borderline high: 200 – 240 mg/dL
  • High risk: > 240 mg/dL
  • Critical/Alert: > 300 mg/dL

12. Laboratory Interpretation

Hypercholesterolemia occurs in hypothyroidism (Myxedema), nephrotic syndrome, atherosclerosis, arteriosclerosis, uncontrolled diabetes, and obstructive jaundice. Hypocholesterolemia is seen in hyperthyroidism and acanthocytosis.

13. Potential Sources of Variability

  • Lysed plasma specimens may yield falsely elevated results.
  • Plasma cholesterol is stable for 7 days at room temperature and 6 months at –20 °C.
  • Do not use turbid reagents; this indicates contamination.

14. References

  1. Richmond W. Clin Chem 1973; 19:1350.
  2. Tarbutton PN, Gunter CR. Clin Chem 1974; 20:724.
  3. Allain CC et al. Clin Chem 1974; 20:470.
  4. Richmond W. Scand J Clin Lab Invest 1972; 29 (Suppl 26): Abst 3.25.
  5. Young DS et al. 1975; 21D.

Monday, October 27, 2025

BILIRUBIN TEST. Information about test and diagnosis of diseases

Dr. Yousef Al Adbai. 
              BILIRUBIN TEST 

1. Purpose: Quantitative estimation of serum bilirubin (Total and Direct) by Jendrassik and Grof Method. Measurement of total bilirubin is useful in the diagnosis of jaundice due to any cause and is an indicator of liver function.

2. Principle: Bilirubin reacts with diazotized sulfanilic acid to form an azo dye which is red in neutral and blue in alkaline solution. Whereas the water-soluble bilirubin glucuronides react “directly” (the free bilirubin). “Indirect” bilirubin reacts only in the presence of an accelerator. The total bilirubin in serum or plasma is determined using by coupling with diazotized sulfanilic acid after the addition of caffeine, sodium benzoate and sodium acetate. A blue azobilirubin is formed in alkaline Fehling solution II. This blue compound can also be determined selectively in the presence of yellow byproducts (green mixed coloration) by photometry at 578 nm. The direct bilirubin is measured as the red azo dye at 546 nm using the method of Schellong and Wende without the addition of alkali.

3. Performance specifications:
3.1. Linearity: Up to 20 mg/dL.
3.2. Measurement range: As low as 0.05 mg in serum.
3.3.  Sensitivity: Lower detection limit is 0.05 mg/dL.

4. Primary sample:
4.1. Use only serum as specimen for the test.
4.2. Collect 4 mL of venous blood in a plain vacutainer tube.
4.3. Do not expose samples for serum bilirubin estimation to tube light/sunlight.
4.4. Do not use hemolyzed, contaminated or lipemic sera.
4.5. Separate serum as soon as possible; Store the serum at −10°C until required, for a maximum up to one month.

5. Type of container and additive: Use plain vacutainer tubes for collecting samples.

6. Reagents/Consumables:
6.1. Sulfanilic acid
6.2. Accelerator: Caffeine, sodium benzoate, sodium acetate
6.3. Sodium nitrite
6.4. Fehling solution II: 930 mmol/L Potassium sodium tartrate, 1.9 mol/L sodium hydroxide solution.

7. Instrument: Semi-autoanalyzer.

8. Procedure:
8.1. Switch on the machine and press “FLUSH” button by keeping the tubing in distilled water for 2 minutes.
8.2. Press “PROC”. Different test procedures will be displayed.
8.3. Select the test to be processed by entering its number and then press “ENTER” key.
8.4. Now the assay parameters of the specific test procedure will be displayed. Note down the volume of the reagent and the sample to be used.
8.5. Feed the reagent blank with each batch of patient samples.
8.6. Then feed the standard followed by test samples and record the values.
   Assay: End point
· Sample volume: 50 µL
· Reagent volume: 500 µL
· No. of readings: 3
· Incubation time: 15 minutes

Total Bilirubin Procedure:
Measure the absorbance of the sample against distilled water or if necessary against the blank.

Direct Bilirubin Procedure:
Measure the absorbance of the sample against distilled water or if necessary against the blank.

9. Interferences: Turbid lipemic and lysed sera. In patients taking heavy dose of B complex, riboflavin and the folate in it may interfere by giving yellow color to the blood and urine.

10. Calculating results: For measurements against a blank: Total bilirubin concentration = A × 10.5 mg/dL.

11. Biological Reference Range:
    *   Total Bilirubin– up to 1.0 mg/dL
    *   Indirect or unconjugated bilirubin– 0.1 to 0.6 mg/dL
    *   Direct or conjugated bilirubin– up to 0.3 mg/dL

12. Critical/Alert level values: 3.0 mg/dL

13. Laboratory interpretation: Increase of bilirubin suggests jaundice; increase of both total and direct bilirubin suggests obstructive/hepato-cellular jaundice. Increase of total bilirubin alone with normal direct bilirubin suggests hemolytic jaundice. A level of 0.4 mg/dL or more of direct bilirubin suggests liver involvement.

14. Potential sources of variability:
    *   Lysed serum specimens may give falsely elevated values.
    *   Dilute the specimen if the bilirubin value is> 10 mg/dL suitable dilution can be done with normal saline. In such a case, the results obtained should be multiplied by dilution factor to be obtained correct bilirubin value.

                  Reference:
 Manual of Medical Laboratory Techniques

CREATININE TEST. Information about test and diagnosis of diseases

Dr. Yousef Al Adbai. 
                    Creatinine test

1. Purpose: Quantitative estimation of creatinine in human plasma by modified Jaffe’s method (initial rate or fixed time method) Measurement of plasma creatinine is useful in the diagnosis, treatment and follow-up of renal diseases/renal failure. Increase of serum creatinine indicates a definite damage of renal tissue.

2. Principle: Creatinine reacts with picric acid in alkaline medium to form an orange-yellow colored complex of creatinine picrate. This colored complex absorbs light at 492 nm the rate of increase in absorbance is directly proportional to the creatinine concentration in the sample.
3. Performance specifications:
3.1. Linearity: Up to 24 mg/dL in plasma.
3.2. Measurement range: 0.1–24 mg/dL of creatinine in plasma.
3.3. Sensitivity: The minimum detection limit by this kit is 1.0 mg/dL.
3.4. Specificity: This method measures a number of other non-creatinine substances also other than creatinine.

4. Primary sample:
4.1. Use only plasma as specimen for the test.
4.2. Collect 4 mL of venous blood in a heparin vacutainer tube.
4.3. Do not use lysed plasma for testing as it may give very high results.
4.4. Do not use contaminated/turbid samples for testing.
4.5. Process the sample on the same day within 3 hours of collection.

5. Type of container and additive: Use heparin vacutainer tubes for collecting blood samples.

6. Instrument: Semi-autoanalyzer

7. Reagents/Consumables:
7.1. Creatinine reagent: Picric acid 8.73 mmol/L
7.2. Buffer solution: 200 mmol/L of sodium hydroxide
7.3. 25 mmol/L of phosphate
7.4. Creatinine standard (2 mg/dL) solution containing creatinine in hydrochloric acid with preservative.

8. Procedure:
8.1. Switch on the machine and press “FLUSH ” button by keeping the tubing in a container with 2% detergent for 2 minutes followed by distilled water for 2 minutes.
8.2. Press “PROC”. Different test procedures will be displayed.
8.3. Select the test to be processed by entering its number and then press “ENTER” key.
8.4. Now the assay parameters of the specific test procedure will be displayed. Note down the volume of the reagent and the sample to be used.
8.5. Feed the blank with each batch and ensure the absorbance of reagent blank to zero
8.6. Then feed the test samples and record the values.
Assay: Fixed time (Initial rate) Reagent volume:500μL/1000μL
Wavelength: 490 nm Sample volume: 25μL/50μL
Temperature: 37°C Zero setting with distilled water
No. of readings: 2 Time: 60 sec
Concentration of Std: 2 mg/dL

9. Calculating results:
Sample absorbance× Concentration of standard = Sample concentration
Standard absorbance

10. Biological reference Range:
Male: 0.7 –1.4 mg/dL
Female: 0.6 –1.2 mg/dL

11. Critical/Mert level values: More than 30 mg/dL.

12. Laboratory interpretation: Increase of creatinine in blood suggests kidney damage.
Example: Chronic glomerulonephritis.

13. Potential sources of variability:
13.1. Lysed plasma specimens may give falsely elevated values
13.2. Creatinine remains stable in plasma for up to 2 days.
13.3. Number of substances other than creatinine interfere with the assay.

               Reference :
 Manual of Medical Laboratory Techniques

Glucose Test. Information about test and diagnosis of diseases

 Dr. Yousef Al Adbai. 
                 Glucose Test

1. Purpose
: Quantitative estimation of glucose in human serum or plasma or Cerebrospinal fluid (CSF) or other body fluids by enzymatic method (GOD-POD). Plasma Glucose determinations are useful in the diagnosis and treatment of diabetes mellitus and in monitoring the response to treatment of diabetes mellitus with insulin or oral hypoglycemic agents. Elevated glucose levels may be associated with pituitary or thyroid dysfunction, renal failure and liver disease, whereas low glucose levels may be associated with insulinoma, hypopituitary neoplasms, or insulin induced hypoglycemia. CSF and fluids have increased glucose in diabetic condition.
2. Principle: Glucose oxidase (GOD) converts glucose to gluconic acid. Hydrogen peroxide formed in this reaction, in the presence of peroxidase (POD), oxidatively couples with 4 - aminoantipyrine (AAP) and phenol to produce red quinone-imine dye. This dye has absorbance maximum at 505 nm. The intensity of color complex is directly proportional to the concentration of glucose in specimen.
3. Performance specifications 
3.1. Linearity: Up to 500 mg/dL of plasma. 
3.2. Measurement range: 40–500 mg/dl
3.3. Sensitivity: The minimum detection limit by this kit is 40 mg/dL
4. Primary sample
4.1. Use only plasma as specimen for the test
4.2. Collect 2 mL of venous blood in a Fluoride—EDTA mixture tube Heparin vacutainer tube.
4.3. Do not use lysed plasma for testing as it may give very high results 
4.4. Do not use contaminated/turbid samples for testing 
Process the sample on the same day within 3 hours of collection.
4.6. Type of container and additive: Fluoride—EDTA mixture tube. 
5. Equipment: Semi-autoanalyzer
6. Reagents: Phosphate buffer; pH 7.5; glucose oxidase; peroxidase; 4 aminoantipyrine; phenol
7. Procedure: 
7.1. Switch on the machine and press “FLUSH “ button by keeping the tubing in a container with 2% detergent for 2 minutes followed by distilled water for 2 minutes
7.2. Press “PROC”. Different test procedures will be displayed. 
7.3. Select the test to be processed by entering its number and then press “ENTER” key.
7.4. Now the assay parameters of the specific test procedure will be displayed. Note down the volume of the reagent and the sample to be used. 
7.5. Feed the blank with each batch and ensure the absorbance of the blank is less than 0.15. If the absorbance of the ‘blank is more than 0.15’ discard the reagent.
7.6. Then feed the test samples and record the values. 
7.7. Check whether the sample is hemolyzed, icteric or lipemic before processing. If the sample is lysed, collect another sample and proceed.
Assay: 
End point 
Reagent volume: 1.0 mL
Wavelength: 505 nm (500–550)
Sample volume: 10 μL
Temperature: 37°C
Zero setting with Reagent blank
Incubation: 5 minutes
8. Interference: Turbid, lipemic, hemolyzed samples, high levels of ascorbic acid, and plasma bilirubin will interfere. Oxalate and fluoride do not interfere.
9. Calculating results:
 10. Biological reference range: 
Glucose Fasting is 60–110 mg/dL. Glucose PP is 90–140 mg/dL Glucose Random is 60–130 mg/dL
11. Critical/Alert level values: 
Below 60 mg/dL 
Above 400 mg/dL
12. Laboratory interpretation: 
Increase of blood glucose usually in diabetes mellitus, decrease in insulinoma. Decrease of CSF sugar in infection. Increase of CSF sugar in hyperglycemia.
13. Potential sources of variability:
13.1. Do not use if the absorbance of the blank reagent is greater than 0.15 at 500 nm as it indicates deterioration of the reagent. 
13.2. Check if the patient has followed the instructions regarding preparation before collecting samples for fasting/post- prandial, plasma glucose/glucose tolerance test.
13.3. The periodic update on the reference ranges needs to be made note of.

BIBLIOGRAPHY
1. Trinder P. Ann Clin Biochem 1969; 6: 24.

Reference➵ Manual of Medical Laboratory Techniques

Sunday, October 26, 2025

Urea test. Information about test and diagnosis of diseases

Dr. Yousef Al Adbai. 
                          UREA

1. Purpose: Quantitative estimation of urea in human serum by Urease-GLDH/UV kinetic method. Determination of serum urea nitrogen is an important index of kidney function. Impaired renal function or increased tissue protein breakdown is associated with increased urea nitrogen levels, whereas liver damage or pregnancy is associated with decreased levels.
2. Principle: Urea is hydrolyzed by urease to form ammonium carbonate. In the second reaction 2-oxoglutarate reacts with ammonium ion in the presence of glutamate dehydrogenase (GLDH) and the coenzyme NADH to produce L-glutamate. In this reaction two moles of NADH are oxidized to NAD^+ for each mole of urea hydrolyzed. The rate of decrease in the NADH concentration is directly proportional to the urea concentration in the specimen. It is determined by measuring the absorbance at 340 nm.
   
3. Performance specifications:
   · 3.1. Linearity: Up to 240 mg/dL of serum
   · 3.2. Measurement range: 2–240 mg/dL
   · 3.3. Sensitivity: Lower limit of detection is 2 mg/dL
4. Primary sample:
   · 4.1. Use plasma
   · 4.2. Collect 2 mL of venous blood from a peripheral vein in a heparin vasculature tube
   · 4.3. Do not use hemolyzed/contaminated plasma for testing
5. Type of container and additive
   · Use heparin/plain vasculature tubes for collecting samples; do not use hemolyzed/contaminated plasma for testing
6. Instrument: Semi-automatizer
7. Reagents/Consumables: The reconstituted reagent contains the following:
   · 7.1. TRIS pH 7.8, 2-Oxoglutarate, ADP, Urease, GLDH
   · 7.2. NADH
   · 7.3. Urea (50 mg/dL)
8. Procedure:
   · 8.1. Switch on the machine and press “FLUSH” button by keeping the tubing in a container with 2% detergent for 2 minutes followed by distilled water for 2 minutes.
· 8.2. Press “PROC”, different test procedures will be displayed.
· 8.3. Select the test to be processed by entering its number and then press “ENTER key.”
· 8.4. Now the assay parameters of the specific test procedure will be displayed. Note down the volume of the reagent and the sample to be used.
· 8.5. Run the standard with each batch of patient sample.
· 8.6. Then feed the test samples and record the values.
  · Assay: 2-point Kinetic
  · Sample volume: 10 μL
  · Wavelength: 340 nm
  · Reagent volume: 400 μL
  · Start reagent 100 μL
  · Temperature: 37 °C

9. Biological reference range: 15–38.5 mg/dL
10. Alert/Critical values: Above 80.0 mg/dL
11. Laboratory interpretation: Increase suggests impaired renal function, acute nephritis, chronic glomerulonephritis.
12. Potential sources of variability
   · 12.1. Use of only clear, unhemolyzed plasma separated from the erythrocytes as soon as possible. Lysed plasma specimens may give falsely elevated values
   · 12.2. On storage, the working reagent may develop a pink color which makes the use of reagent blanks necessary with every run.
   · 12.3. This method is recommended to perform only on mechanized equipment. It is difficult to include all samples and reagent blank exactly for the same intervals.
   · 12.4. The scheme may use for adaptation purpose for instruments with no specific adaptation sheet.

BIBLIOGRAPHY

1. Kassiter JP. New Eng J Med 1971;285:385.
2. Mackay EM, Mackay LL. Clin.Invest 1927;4:295.
3. Talke HN, Schubert, GE Kin. Wschr 1965;42:174.

Glucose tolerance Test. Information about test and diagnosis of diseases

Dr. Yousef Al Adbai. 


GLUCOSE TOLERANCE TEST

1. Purpose:

To recognize milder cases of diabetes and renal glycosuria. Quantitative estimation of glucose in human plasma by enzymatic method (GOD-POD). Plasma glucose determinations are useful in the diagnosis and treatment of diabetes mellitus and in monitoring the response to treatment of diabetes mellitus with insulin or oral hypoglycemic agents. Elevated glucose levels may be associated with pituitary or thyroid dysfunction, renal failure, and liver disease, whereas low glucose levels may be associated with insulinoma, hypopituitarism, or insulin-induced hypoglycemia. CSF and fluids have increased glucose in diabetic condition.

2. Principle:

Glucose oxidase (GOD) converts glucose to gluconic acid. Hydrogen peroxide formed in this reaction in the presence of peroxidase (POD) oxidatively couples with 4-aminoantipyrine and phenol to produce red quinoneimine dye. This dye has absorbance maximum at 505 nm. The intensity of color complex is directly proportional to the concentration of glucose in specimen.

3. Performance specifications:

3.1. Linearity: Up to 600 mg/dL of plasma.
3.2. Measurement range: 25–600 mg/dL.
3.3. Sensitivity: The minimum detection limit 40 mg/dL.

4. Primary sample:

4.1. Use only plasma as specimen for the test.
4.2. Collect 2 mL of venous blood in a fluoride–oxalate mixture tube.
4.3. Do not use lysed plasma for testing as it may give false results.
4.4. Do not use contaminated/turbid samples for testing.
4.5. Process the sample on the same day within 3 hours of collection.

5. Type of container and additive:

Fluoride—EDTA mixture tube.

6. Reagents/Consumables:

For patient use, commercially available glucose (75 g) mixed with water.

7. Instrument:

Semi-autoanalyzer.

8. Procedures:

Instructions to be given to the patient: These instructions should be given to the patients by previous day of the investigation.


Biochemistry

8.1. The patient should not take any food after 9 pm the previous night till the test is performed.
8.2. The subject should have normal diet for at least 3 days prior to the test.
8.3. He/she should not have taken drugs which affect blood sugar.
8.4. In exceptional cases, when the patient has to come from a distant place, light tea without sugar may be allowed (2 hours before collection).

Method: Upon arrival of the patient, the following should be done:
8.5. Body weight should be noted down.
8.6. Fasting blood sample should be collected and glucose estimation should be performed.
8.7. Specimen of fasting urine is collected and test for glucose, albumin, and acetone to be done.
8.8. 75 g of glucose dissolved in 300 mL of water should be given orally.
8.9. Blood and urine samples will be collected for every half an hour interval for 2 hours after the glucose has been taken.
8.10. It is not always possible to collect urine at every half hour interval. In such cases, urine sample can be collected for every 1 hour interval.

Glucose estimation: As per the method given in this manual.
Urine sugar: As per the method given in this manual, with the standard curves.

Normal responses:
Fasting glucose within normal limit. Maximum blood glucose is reached either half or one hour after taking the glucose. The blood glucose then returns rapidly to the normal fasting limits, which are often reached in one and a half hour and almost always at two hours. There should be no sugar in any of the urine specimens.

9. Reference:

The GTT curve will be interpreted with the standard curves.

10. Critical/Alert level values:

Below 40 mg/dL, above 400 mg/dL.

11. Potential sources of variability:

11.1. Do not use if the absorbance of the blank reagent is greater than 0.150 at 500 nm as it indicates deterioration of the reagent.
11.2. Check if the patient has followed the instructions regarding preparation before collecting samples.


BIBLIOGRAPHY:

  1. Harold Varley. Practical Clinical Biochemistry, 5th ed., 1980;1:406–10.
  2. Trinder P. Ann Clin Biochem 1969; 6: 24.

Tuesday, October 21, 2025

The Hidden World Inside a Drop of Blood

🧠 Title: The Hidden World Inside a Drop of Blood

Have you ever looked at a drop of blood and wondered what secrets it holds?
To the naked eye, it’s just red liquid—but under a microscope, it’s a living universe.

Inside that single drop, millions of red blood cells rush to carry oxygen, while white blood cells act like tiny soldiers, protecting your body from invisible invaders.
And then, the platelets—the quiet heroes—stand ready to stop bleeding the moment you get a cut.

Every time you take a blood test, the lab reveals stories your body can’t tell on its own.
From your sugar levels to liver function, each result whispers a clue about your health.

So next time you see a drop of blood, remember — it’s not just blood.
It’s a microscopic city working non-stop to keep you alive. 💉✨




CERULOPLASMIN Test. lab information about test and diagnosis of diseases

______________________________________________________________________________ Ceruloplasmin Test. Lab information about test and diagnosis ...